81001 Universal Signal Averaging Resistivity Meter GTR-2
The new resistivity meter GTR-2 combines an excellent depth penetration with a relatively low output current. It combines a
numerical stacking of consecutive readings with automatic current reversal to produce a high accuracy output.
The fast sampling and continuous calculation of the running average
performed by a sophisticated microprocessor allow virtual signal enhancement. This technique is generally recognised to be the best method to eliminate SP and other noises. The GTR-2
also controls the SP level within each measurement cycle in order to take into account possible variations occuring during the measurement.
Thanks to a highcut filter and to cycle
periods calculated to reject industrial frequencies, the GTR-2 can extract coherent signals from an apparently random background.
Specifications
Voltage measurements đV and SP:
4 digits with automatic ranges Measuring range: ± 5
V
Resolution: 0,01 mV
Inputprotection: ± 600 VDC
Input impedance: 10 Mohm Overall accuracy: ± 2 %
Current measurement :
Measuring range: 5 A max.
Resolution: 0,1 mA Programmable K factor: from 1 to 99.99
Internal batteries: 6 V, 6 Ah rechargeable Ambient temperature: - 5°C to + 50° C Dimensions: 360 x 260 x 155 mm
Weight: 4.5 kg
Number of programmable measuring
cycles: 1 to 32
Duration of programmable cycles: 0,96 s to 5,12 s
All accessories for geotechnical surveying such as electrodes and cables are available from our stock.
|
|
 |
 |
 |
|
Display
As illustrated in the example below, the operator has permanent control over the measurement parameters, including the
apparent resistivity calculated for a K factor previously entered through the keyboard. The operator can follow the evolution of these parameters during the measurement
which can be completed immediately when results obtained are satisfactory.
Memory / RS-232 serial interface
Every parameter displayed can be saved using automatic sequential numbering after
each measurement is completed. Memory size: 1000 measurements.
Data transfer capability to a PC computer via RS232 serial interface is included.
Simplicity of operation
After a few minutes, you’ll become familiar with the instrument. Choose the mode and introduce
the K factor, then press the ‘’MEASURE’’ key. Write the measurement in a notebook or enter it into the memory.
|
|
|
|
|